PRODUCTS
ELECTRICAL PROBE SYSTEMS
THERMAL PLATES
LIQUID CRYSTAL SYSTEMS
RESOURCES
SUPPORT
ABOUT
Flagship mini electrical probe stage
-190°C - 600°C with liquid nitrogen cooling option
Compact design to fit on Raman, optical microscope, reflection FTIR microscope
28 mm × 30 mm sample area
4 independent electrical probes with manual positioning
Optional triaxial connection for pA measurement
Optional extra feedthrough
Temperature Control |
mK2000 with programmable precision LVDC PID method to minimize electrical noise |
Thermal Block |
Silver plate for superior thermal uniformity |
Temperature Minimum |
-190°C (with optional liquid N2 cooling) |
Temperature Maximum |
600°C |
Temperature Sensor |
100 Ω Platinum RTD |
Maximum Heating Rate |
+150°C per minute at 100°C |
Maximum Cooling Rate |
-50°C per minute at 100°C |
Minimum Heating and Cooling Rate |
±0.01°C per minute |
Temperature Resolution |
0.01°C |
Temperature Stability |
±0.05°C (>25°C),±0.1°C (<25°C) |
Power Supply |
Universal power input |
Software |
Windows software to record and export temperature-time data |
Electrical Probes |
DC electrical probe options in tungsten-rhenium, beryllium-copper, or other materials |
Probe Positioning |
Manual cantilever probes independently positionable to any point of the sample area before sealing chamber |
Connectors |
Coaxial, floating, or triaxial BNC connectors Optional extra feedthrough |
Sample Surface |
Grounded, floating, or triaxial options |
Non-magnetic Construction |
Optional construction with non-magnetic materials for magnetically sensitive samples |
Optical Access |
Optical access for reflected light measurements Options for transmitted light measurements are available |
Optical Windows |
Removable and exchangeable windows permit full-spectrum transparency |
Minimum Objective Working Distance |
8.5 mm |
Top Aperture |
Ø18 mm (standard), with Ø22mm window Ø28 mm (optional), with Ø31.75mm window Ø38mm (optional), with Ø42mm window |
Top Viewing Angle |
±47° (standard), ±60.7°(optional) |
Window Defrost |
Integrated external window defrost |
Sample Area |
28 mm x 30 mm (50mm x 50mm option available, contact us about the HCP402SV-PM) |
Chamber Height |
6.3 mm |
Sample Access |
Quick-access cover with spring-loaded screws |
Atmosphere Control |
Gas purging capability to control humidity, condensation, and oxidation |
Frame Cooling |
Integrated frame cooling with optional chiller system |
Mounting |
Horizontal and vertical mounting capability |
Frame Dimensions |
180 mm x 130 mm x 26 mm |
Weight |
1500g light-weight aluminum body |
Active Sample Cooling | |
- None (Standard) | - LN2-P4D3 cooling sytsem |
Window aperture | |
- 18mm (22mm glass) (Standard) | - 38mm (42mm glass) |
Window Material | |
- Quartz Glass (Standard) | - BaF2 Glass for IR |
- Soda Lime Glass | - ZnSe Glass for IR |
- CaF2 Glass for IR | - Saphire Glass for UV/IR |
Sample Mounting | |
- Threaded Holes (Standard) | - Add set of 2 sample clips |
Inner Sample Cover | |
- None (Standard) | - Add inner cover |
Enviromental Control Ports | |
- Straight quick-connect ports (Standard) | - SS bellows with KF-16 flange |
Mounting Adapters | |
- L-Brackets (Standard) | - Vertical |
- Horizontal |
- Vertical with rotation |
Electrically Biased Sample Area | |
- Grounded (Standard) | - Floating with Coax feedthrough |
Temperature Controller | |
- mK2000B with switching PID output (Standard) |
- mK2000B with LVDC output upgrade |
Upgrade to HCP621G-PM+ or HCS621G-PM+ for full customization options. Customizations include:
- KBR, Kapton or Teflon Windows | - Custom lids |
- Temperature limit increase | - Angled windows |
- Custom sample holders |
... And much more! |
The Standard options above can be chosen during purchasing without the need for a custom design. Pricing varies per option, and per combination of options, so contact sales@instec.com for more information.